RF Probes


RF Current Monitoring Probes

The voltage produced by RF current monitoring probes into a 50-ohm load is proportional to the current flowing through the wires going through the probe aperture. RF current monitoring probes, like LISNs, are used to measure conducted emissions. RF current probes measure the current of conducted emissions, whereas LISNs measure the voltage. Current monitoring probes are not intrusive. It is not necessary to separate the wire harness under test. In addition, RF current monitoring probes typically cover a wider frequency range than LISNs. The EMC standards indicate where LISNs must be used and where current monitoring probes must be used. The primary use is conducted emission measurement on data or control cables. Furthermore, RF current monitoring probes are very versatile tools for the investigation and improvement of EMC issues and for many other applications.
Tekbox RF current monitoring probes cover the frequency range up to 1 GHz. The series with fixed aperture are especially inexpensive. In circumstances, where the cable harness cannot simply be disconnected to feed it through the aperture, Tekbox offers the TBCP2 series snap on probes.
To calculate the corresponding current flowing through the wires passing the probe aperture, subtract the individually documented trans-impedance value in [dB] from the probe’s output voltage.

ModelTBCP1-200TBCP1-250TBCP1-500TBCP2-250TBCP2-30K400TBCP2-500TBCP2-750TBCP3-1000TBCP4-250TBCP4-500TBCP4-750
Characterized frequency range1 kHz-500 MHz1 kHz-300 MHz1 kHz-500 MHz1 kHz-400 MHz10 Hz-500 MHz1 kHz-600 MHz1 kHz-1 GHz1 kHz-1 GHz1 kHz-400 MHz1 kHz-600 MHz1 kHz-1 GHz
3dB bandwidth40 kHz-200 MHz300 kHz-250 MHz400 kHz-500 MHz300 kHz-300 MHz30 kHz-400 MHz1 MHz-500 MHz1.5 MHz-850 MHz3 MHz-1 GHz200 kHz-300 MHz400 kHz-500 MHz1.5 MHz-850 MHz
Trans-impedance17 dBΩ18 dBΩ19 dBΩ13 dBΩ3 dBΩ16 dBΩ20 dBΩ20 dBΩ13 dBΩ16 dBΩ20 dBΩ
Aperture diameter25 mm25 mm25 mm32 mm32 mm32 mm32 mm17 mm32 mm32 mm32 mm
Snap on / hingenononoyesyesyesyesnononono
Product PageTBCP1-200TBCP1-250TBCP1-500TBCP2-250TBCP2-30K400TBCP2-500TBCP2-750TBCP3-1000TBCP4-250TBCP4-500TBCP4-750

High Saturation RF Current Monitoring Probes

ModelTBCP2-1M800TBCP2-1M100TBCP2-6M300TBCP2-3M900TBCP2-20M700TBCP2-30M1000TBCP5-150K400
Characterized frequency range10 kHz – 1 GHz10 kHz – 200 MHz10 kHz – 500 MHz10 kHz – 1 GHz10 kHz – 800 MHz10 kHz – 1 GHz10 Hz – 500 MHz
3dB bandwidth1 MHz – 800 MHz1 MHz – 100 MHz6 MHz – 300 MHz3 MHz – 900 MHz20 MHz – 700 MHz30 MHz – 900 MHz150 kHz – 400 MHz
Trans-impedance0 dBΩ, typ.3 dBΩ, typ.10 dBΩ, typ.10 dBΩ, typ.20 dBΩ, typ.22 dBΩ ± 2 dBΩ
from 25 MHz – 1 GHz
11 dBΩ
Max. RF Current10 A40 A20 A3 A15 A10 A2 A
Saturation DC-400Hz≥ 400A≥ 400A≥ 400A≥ 100A≥ 100A≥ 50A≥ 200A
Aperture diameter32 mm32 mm32 mm32 mm32 mm32 mm46 mm
Snap on / hingeyesyesyesyesyesyesyes
ApplicationHigh frequency;
High saturation
High frequency;
High saturation
High frequency;
High saturation
High frequency;
High saturation
High frequency;
High saturation
High frequency;
High saturation
High frequency;
High saturation
Product PageTBCP2-1M800TBCP2-1M100TBCP2-6M300TBCP2-3M900TBCP2-20M700TBCP2-30M1000TBCP5-150K400

RF Surface Current Monitoring Probes

RF Surface current monitoring probes are utilised in applications where RF currents flowing on surfaces such as PC groundplanes or traces, metal planes or wires must be measured.

ModelTBSCP1-5M300TBSCP1-10M500
Characterized frequency range30 kHz – 400 MHz30 kHz – 600 MHz
Trans-impedance-7 dBΩ-5 dBΩ
Footprint40 x 15 mm40 x 15 mm
Product PageTBSCP1-5M300TBSCP1-10M500

Coaxial RF Current Monitoring Probes

Coaxial RF current monitoring probes are used to measure RF currents in coaxial cables. They can be used as general purpose coupling devices to measure RF power or as loop antenna transducers.

ModelTBCCP1-2K70TBCCP1-3K100TBCCP1-400K600
Characterized frequency range10 kHz – 100 MHz10 Hz – 100 MHz10 Hz – 600 MHz
Trans-impedance0 dBΩ+5 dBΩ+23 dBΩ
ApplicationTransducer for large loop antennas (LLA)General purpose coupling deviceGeneral purpose coupling device
Product PageTBCCP1-2K70TBCCP1-3K100TBCCP1-400K600

RF pulse Current Monitoring Probes

RF pulse Current Monitoring Probes are typically used for surge or RF pulse current monitoring applications in the time domain.

ModelTBPCP1-20100TBPCP2-3070TBPCP3-05H30TBPCP3-2H40TBPCP3-20H70TBPCP3-100H80TBPCP4-1H30TBPCP4-5H50TBPCP4-50H50TBPCP4-100H60
Trans-impedance @ 50Ω-20 dBΩ
0.1 V/A
-26 dBΩ
0.05 V/A
-26 dBΩ
0.05 V/A
-20 dBΩ
0.1 V/A
-12 dBΩ
0.25 V/A
-6 dBΩ
0.5 V/A
-26 dBΩ
0.05 V/A
-20 dBΩ
0.1 V/A
-12 dBΩ
0.25 V/A
-6 dBΩ
0.5 V/A
Trans-impedance @ HighZ0.2 V/A0.1 V/A0.1V/A0.2V/A0.5V/A1V/A0.1V/A0.2V/A0.5V/A1V/A
3 dB bandwidth20 Hz-100 MHz30 Hz-70 MHz0.5 Hz–30 MHz2 Hz–40 MHz20 Hz–70 MHz100 Hz–80 MHz1 Hz–30 MHz5 Hz–50 MHz50 Hz–50 MHz130 Hz–60 MHz
Characterized frequency range1 Hz-200 MHz1 Hz-200 MHz0.1 Hz–100 MHz1 Hz–100 MHz1 Hz–100 MHz1 Hz–100 MHz0.1 Hz–100 MHz1 Hz–100 MHz1 Hz–100 MHz1 Hz–100 MHz
Droop rate10 %/ms20 %/ms0.12 %/ms0.3 %/ms6 %/ms0.02 %/µs0.2 %/ms0.5 %/ms7 %/ms0.06 %/ms
Rise time< 5 ns< 5 ns< 10 ns< 9 ns< 9 ns< 8 ns< 10 ns< 9 ns< 10 ns< 10 ns
Max. peak current3000 A600 A5000 A2000 A1500 A600 A5000 A3500 A2500 A1000 A
IT Max0.08 As0.02 As0.25 As0.1 As0.03 As0.007 As0.7 As0.4 As0.1 As0.02 As
I/f0.3 peak A/Hz0.07 peak A/Hz1.5 peak A/Hz0.6 peak A/Hz0.125 peak A/Hz0.06 peak A/Hz2.8 peak A/Hz1.5 peak A/Hz0.5 peak A/Hz0.25 peak A/Hz
Max. RMS current120 A15 A55 A70 A40 A20 A150 A150 A60 A60 A
Aperture diameter25 mm5.5 mm15 mm15 mm15 mm15 mm46 mm46 mm46 mm46 mm
Outer diameter76 mm27 mm61 mm61 mm61 mm61 mm124 mm124 mm124 mm124 mm
Height31 mm19 mm30 mm30 mm30 mm30 mm37 mm37 mm37 mm37 mm
Weight320 g60 g320 g320 g320 g320 g1050 g1050 g1050 g1050 g
Product PageTBPCP1-20100TBPCP2-3070TBPCP3-05H30TBPCP3-2H40TBPCP3-20H70TBPCP3-100H80TBPCP4-1H30TBPCP4-5H50TBPCP4-50H50TBPCP4-100H60

Current Probe calibration fixtures

Current Probe Calibration Fixtures are used for calibrating / measuring the transimpedance of current probes. Calibrators are also required for the calibration of immunity test setups with BCI probes.

ModelTBCP1-CALTBCP2-CALTBCP3-CALTBCP4-CALTBCP5-CALTBPCP2-CALTBPCP3-CALTBPCP4-CALTBEMCL1-CAL
ApplicationTBCP1 series, TBPCP1-20100TBCP2 seriesTBCP3 seriesTBCP4 seriesTBCP5 seriesTBPCP2-3070TBPCP3 seriesTBPCP4 seriesEM-clamp calibrator
Product PageTBCP1 series, TBPCP1-20100TBCP2 seriesTBCP3 seriesTBCP4 seriesTBEMCL1-CAL

Bulk Current Injection Probes

Bulk Current Injection Probes are used to inject RF-current into cables of electrical equipment to test the susceptibility against radiated electromagnetic energy. Tekbox BCI-probes can be clamped around wire bundles with up to 27 mm. They are designed to meet the specifcations of the standards IC 61000-4-6 (domestic and industrial products) and ISO 11452-4 (automotive products).

Tekbox supplies suitable wideband power amplifiers to drive the BCI probes:

TBMDA-CDN25 for immunity tests according to IC 61000-4-6
TBMDA-BCI25 for immunity tests according to ISO 11452-4

The calibration fixture TBBCI1-CAL is suitable for both BCI-probe models

ModelTBBCI1-200K280TBBCI1-300K420-HVTBBCI1-800K420TBBCI1-CAL
Frequency range150 kHz – 230 MHz300 kHz – 400 MHz1 MHz – 400 MHz150 kHz – 400 MHz
Coupling loss8 dB typical5 dB typical5 dB typical
Maximum RF input power50W continuous
100W short time
50W continuous
100W short time
50W continuous
100W short time
500 W
Aperture diameter27 mm27 mm27 mm
Outer dimensions92 x 76 mm92 x 76 mm92 x 76 mm200 × 150 × 140 mm
Application / StandardBCI probe for IEC / ISO 61000-6-4BCI probe for ISO 11452-4; MIL-STD-461 CS114, GJB 151B-2013, DO-160BCI probe for IEC / ISO 11452-4BCI probe calibrator for IC / ISO 61000-6-4
IEC / ISO 11452-4
Product PageTBBCI1-200K280TBBCI1-300K420-HVTBBCI1-800K420TBBCI1-CAL

Current Coupler

ModelTBMHA1
Mains Voltage Range100 V–120 V/200 V–240 V switchable
Max Mains Current16A, depending on country specific output socket
Transducer factor50 mV/A with 50 Ohm load resistor
Mains Input SocketC19, 16 A
Mains Output Socketcountry specific
Current Output SocketBNC female
Product PageTBMHA1

Powerline Ripple Detectors

ModelTBPRD-CS101-240V
Characterized frequency range10 Hz to 300 kHz
Insertion loss above 50 Hz>80 dB ± 1 dB with toggle
in "Flat response" position
Input voltage range100 V AC to 240 V AC
100 V DC to 270 V DC
Maximum input voltage240 V AC, 270 V DC
Input connectors4 mm safety sockets
Output connectorBNC female
Fuse50 mA, fast
Littlefuse 0217.050MXP
Product PageTBPRD-CS101-240V